Skip To Content
Glossary

De Novo Submission

What Is a De Novo Submission?

A De Novo classification request is an FDA pathway for certain novel medical devices that have no legally marketed predicate but present low to moderate risk. It allows the FDA to classify a new device type into Class I or Class II when general controls, or general and special controls together, can provide reasonable assurance of safety and effectiveness.

If the FDA grants a De Novo request, the device receives marketing authorization and the agency establishes a classification for the device type. This is not called a 510(k) clearance or a PMA approval. The newly classified device may also become a predicate for future 510(k) submissions when appropriate. See the FDA’s De Novo Classification Request overview.

When Is a Device Eligible for the De Novo Pathway?

De Novo may be appropriate when a device has no suitable legally marketed predicate and its risks can be adequately controlled without the PMA pathway typically associated with Class III devices. A company can generally request De Novo classification directly or after receiving a not substantially equivalent (NSE) determination on a 510(k).

Eligibility is not based on novelty alone. Sponsors must explain the device’s intended use, its risks and benefits, why existing device types are not suitable predicates, and how the proposed controls address relevant risks. Early regulatory planning helps avoid spending time preparing evidence for the wrong pathway. QuickVault’s MedTech Regulatory Submissions Made Easy webinar offers a useful comparison of common pathways.

FDA De Novo Submission Requirements

A De Novo request should make a coherent case for a new device classification and the controls needed to support it. Depending on device characteristics, FDA may expect:

  • Device description: Intended use, indications, design, materials, technology, operating principles, accessories, and labeling.
  • Regulatory rationale: Why no suitable predicate exists, the proposed classification, and the rationale for general or special controls.
  • Benefit-risk analysis: Identified hazards, risk controls, residual risks, probable benefits, and how the evidence supports acceptable risk.
  • Performance evidence: Relevant bench testing, verification and validation, software, usability, cybersecurity, biocompatibility, or other studies; clinical evidence when needed to evaluate safety and effectiveness.
  • Proposed special controls: Device-specific requirements that, with general controls, would provide reasonable assurance of safety and effectiveness for a Class II device.
  • Traceable supporting records: Applicable design controls, risk management, controlled test reports, and documented decisions.

The exact content depends on the device. FDA’s De Novo request content requirements provide the regulatory foundation; device-specific guidance and any FDA feedback should inform the evidence plan.

What Is the FDA De Novo Submission Process?

  1. Confirm pathway eligibility. Research existing classifications and predicates. Determine whether direct De Novo, 510(k), PMA, or another pathway is appropriate.
  2. Define intended use and risks. Establish clear indications for use, identify hazards, and propose a risk-based classification and appropriate controls.
  3. Develop the evidence. Complete the relevant design work, risk assessments, verification, validation, performance testing, and clinical evaluation where needed. Maintain document control over the records.
  4. Prepare the De Novo request in eSTAR. Since October 1, 2025, De Novo requests must use the FDA’s eSTAR electronic submission template unless exempted. The template structures the submission but does not replace the underlying regulatory analysis.
  5. Submit and address FDA questions. The FDA evaluates eligibility, the adequacy of evidence, and whether the proposed controls can reasonably assure safety and effectiveness. Additional information requests may extend the overall timeline.
  6. Receive a grant or decline decision. If granted, FDA authorizes marketing subject to applicable requirements and creates a new classification regulation. If declined, the device cannot be marketed under the requested De Novo authorization.

The FDA’s decision goal for De Novo requests is generally 150 FDA review days, excluding time a request is on hold. That is a review performance goal, not a guaranteed calendar-day launch date.

Consideration

De Novo

510(k) Premarket Notification

Typical Use

Novel low- to moderate-risk device without a suitable predicate

Device that can demonstrate substantial equivalence to a predicate

Core FDA evaluation

Risk-based classification and adequacy of controls

Substantial equivalence

Predicate needed

No suitable predicate available

Generally yes

Outcome

Granted classification request and marketing authorization

510(k) clearance

Future impact

Can establish a new device type that may serve as a predicate

Relies on an established device type or suitable predicate

A De Novo grant may make a later 510(k) pathway possible for similar devices. It does not mean every later device of that type will automatically qualify for clearance. Learn more from FDA’s 510(k) Premarket Notification guidance.

Common De Novo Challenges and Best Practices

De Novo submissions require more than showing a device is innovative. The work is in proving that the proposed controls and evidence match the risks.

  • Unclear eligibility: Document predicate searches and the rationale for the pathway before committing to a submission plan.
  • Weak benefit-risk reasoning: Connect each significant hazard to mitigation measures, testing, labeling, and residual-risk conclusions. Use ISO 14971-related risk practices where applicable.
  • Unconvincing special controls: Make proposed controls specific enough to be meaningful and support them with data. A broad statement that risks are “managed” is not a substitute for evidence.
  • Scattered documentation: Maintain traceability between design inputs, tests, risk evaluations, and the final request; reconcile changes before submission.
  • Late cross-functional review: Have Regulatory, Quality, Engineering, Clinical, and other relevant experts review the same controlled evidence and prepare owners for FDA questions.

After authorization, continue monitoring device performance and maintaining applicable quality-system, complaint, and postmarket processes. An FDA grant is a regulatory milestone, not the end of lifecycle obligations.

How QuickVault Supports De Novo Submission Readiness

QuickVault by Veeva brings design documentation, risk records, controlled documents, and regulatory submission workflows into one environment. Teams can connect supporting evidence to their submission, coordinate review tasks, track questions from regulators, and maintain links to the underlying product information as it evolves.

An eQMS does not determine De Novo eligibility or replace regulatory judgment. Its role is to make the rationale, evidence, and follow-up work easier to organize and review. Explore QuickVault’s regulatory submission management software and its approach to connected design and risk management.

Frequently Asked Questions About De Novo

Is De Novo the same as 510(k) clearance?

No. FDA grants a De Novo classification request and authorizes marketing when the request meets the criteria. A 510(k) results in clearance based on substantial equivalence.

Not necessarily. FDA permits a direct De Novo request when the eligibility criteria are met; an NSE determination from a prior 510(k) is another potential route.

No. The evidence is device-specific. Nonclinical testing may be enough for some devices, while others need clinical evidence to evaluate benefits, risks, and appropriate controls.

Yes, unless exempted. The requirement took effect on October 1, 2025. Review FDA’s eSTAR instructions for the current requirements.

The authorized device may be marketed in accordance with applicable controls. The new classification may also support future 510(k) submissions for appropriate devices, while the manufacturer continues to meet relevant quality and postmarket requirements.